LCR Meter, with D-display
built-in Limit Comparator
@Data sheet(PDF)----- @@ @@
@@@@@Determination display of [GO][NG] by digital comparator inclusion
@
@@ œMeasuring range :
@ L : 0.1ƒΚH ` 199.9H
@ C : 0.1pF ` 1999ƒΚF
@ R : 1mƒΆ ` 1999kƒΆ
@ D : 0.001 ` 1.999
œL,C,R 3 1/2digit, D 3 1/2digit, LED display
œSampling time : 10 times per second approx.
@ (Auto range changeover time:1 step 100 milli second)
œL,C,R and d, analog voltage output
@ (full scale DC2V)
@
‘THE OUTLINE‘
The Digital LCR Meter, Model AX-222 offers easy measurement of contact resistance of electric
components such as a coil, condenser, resistor, switch, relay etc., of inner resistance of a battery,
of junction capacity of semiconductor and of any kind of elements. Determination display of
[GO][NG] by digital comparator inclusion.
Auto range and auto mode functions select automatically an optimum range for the measuring
object with unknown value. Also, manual range changeover is possible and thus unnecessary
time for range changeover can be eliminated.
The display unit displays LCR value in 3-1/2 digit and in case of the measurement of L or C,
the value of D(loss coefficient) is also displayed in 3-1/2 digit.
Analog voltage proportional to the measured value is output, too, for the connection to an analog
recorder, analog comparator, etc.
@@
‘SPECIFICATIONS‘
Measuring range and Accuracy (at23Ž}5Ž)
Measuring range and Accuracy (at23Ž}5Ž)
@L@ 0.1ƒΚH`199.9H }(0.3 of rdg + 2digit)TYP
@C@ 0.1pF`1999ƒΚF }(0.3 of rdg + 2digit)TYP
@R@ 1mƒΆ`1999kƒΆ }(0.3 of rdg + 2digit)TYP
@D@ 0.001`1.999 }(0.5 of rdg + 2digit)TYP
Measuring frequency 1kHz }5%
Measuring signal level Differs according to the range. L:10mA`1ƒΚA C:1V`0.01V
R:100mA`1ƒΚA
Range changeover Auto or manual
Display L,C,R 3-1/2 digit, D 3-1/2 digit, LED display
Sampling time 10 times per second approx.
(Auto range changeover time:1 step 100 milli second)
Changeover of measuring item L and D, C and D, R push-button selectable
Measuring mode
@L@ Series equivalent circuit
@C@ 200pF, 2nF, 20nF, 200nF range(parallel equivalent circuit)
2ƒΚF, 20ƒΚF, 200ƒΚF, 2000ƒΚF range(series equivalent circuit)
@R@ Resistance, parallel resistance in inductance measurement
Measuring terminal 5 terminals
Output signal L, C, R and D, analog voltage output(full scale DC2V)
External bias 0 to 50V DC, 0.1A max.(capacitance measurement)
Comparator Limit L/C/R : 1999 Dual
D : 1999Single
GO/NO-GO Output LED lamp buzzer
SPDT relay output
Power supply AC100V/117V/220V/240V }10%, 50/60Hz, 30VA approx.
Outer dimension 330(W)~85(H)~245(D)mm
(excluding protruding parts such as rubber legs, etc.)
Weight about 3kg
Option BCD Parallel Data out