Low Test Current DC Low-OHM Meter
@Data sheet(PDF)----- @@ @@
@@ Circuit designed to eliminate contact error as
@ much as possible
Most suitable to measure super-low resistance
@ shunt-resistor, etc.
Measurement not to influenced by thermal
@ electro motive force.
Impressed the measuring current at pulse interval,
@in order to reduce abrasion of measuring terminal.
Contact-check is equipped as standard function.
Measuring range : 0.00m`15.000k
Available to measure the persent the set reference
@ value : 1m`10k }50.00%
@ [Minimum resolution of 1Ohm]
GP-IB/RS-232C/Centronics output available.[Option]
Comparison result by built-in comparator is
@ open-collector output and displayed as LO, GO, HI by
@ LED and buzzer.

Measuring current/voltage-Check are built-in as
@ standard function.
THE OUTLINE
AX-1155B is available to measure the super-low resistance with high speed and high accuracy from 0.00m
to 15.000(% measurement : 1m`10k)
In addiition, it is possible tom make the measurement the measuring object accumulately and stable,
because they have DOUBLE MEASURING MODE and special AUTO-ZERO circuit that it is available to
measure them after cancelation of thermoelectromotive force for error factor when the measurement is
made low-resistance especially.
This checker is displayed the measured value, judged HI, GO, and LO, and then the signal is output.
The measuring speed is changeable and it is selectable AVERAGE/DOUBLE/SLOW etc.
And as CONTACT CHECK CIRCUIT and measuring current/abnormal voltage check circuit are equiped as
standard, this checker become higher the confidence.
And it is available to equip the interface of GP-IB, RS-232C, or Centro Output as optional function.
@@
SPECIFICATIONS
Measuring range and Accuracy (at23}5)
Range Measuring range Resolution Accuracy [Slow]
100m 0.00m`150.00m 10ʃ within
}0.02%rdg
}3digit[Average]
}4digit[Slow]
}5digit[Fast]
1 0.0000`1.5000m 100ʃ
10 0.000`15.000 1m
100 0.00`150.00 10m
1k 0.0000k`1.5000k 100m
10k 0`15000 100A
% 1m`9.99m}50.0% -Ditto- Within[(Range/Standard)/5]~0.01%}digit
10m`10k}50.00% Within[(Range/Standard)/5]~0.01%}digit
*:Average=1 Slow=2 Fast=3 :Average=3 Slow=4 Fast=5
Measuring method 4-terminal measurement with contact check
Sampling time Free running mode : 2`10 times/sec.
Remote start about 15msec./30msec.
[Single Mode](Difference by mode.)
Comparator range Ohm mode(0`15000) or
% mode(}50.00) both for the upper and lower limit.
Indication of comparator's comparison result LO/GO/HI LED display and buzzer
Control signal remote start input
remote hold input
comarison output[LO/GO/HI]
contact error output
end of comparison output
Operation condition [Temp.] +5`+40@@[Humidity] below 85%
Power supply AC100V/117V/220V/240V switchable, 50/60Hz, below 50VA
Outer dimension 333(W)~99(H)~300(D)mm
(excluding protruding parts such as rubber legs, etc.)
Weight about 4kg
@@
OPTION
GP-IB Interface/RS-232C Interface/Printer output(8 bit parallel Centronics)
/Connecting cable to printer
*Note:Only one kind of interface can be provided with the checker.